This calibration procedure applies to all models (34460A/61A/65A/70A).
Review Test Considerations and Gain Calibration Considerations before beginning this procedure.
Configuration: Frequency, 10 V range
![]() |
The frequency accuracy of the Fluke 5720A is insufficient to calibrate the DMM. Its frequency output needs to be calibrated against a more accurate reference. The Keysight 33500B Series Function/Arbitrary Waveform Generator is recommended for this calibration procedure. |
![]() |
This calibration takes approximately 1 second to complete. |
Next Procedure |
![]() |